Detail of Recall

Type of Product:Device
Medsafe Reference:24622
Brand Name:Elecsys Syphilis
Model:Cobas e 411/601/602
Affected:
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:Roche Diagnostics GmbH
Issue:The manufacturer has identified that erroneous values were stated in the method sheet (version 1.0 and 2.0) of Elecsys Syphilis, 200 test/kit on cobas e 411, cobas e 601 and cobas e 602, section limitations – interference where the potential interference of IgG and IgM is listed.
In the column Concentration tested 32 g/dL is written instead of the correct 3.2 g/dL for
IgG, and 10 g/dL is written instead of the correct 1.0 g/dL for IgM.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/5/2019