Detail of Recall

Type of Product:Device
Medsafe Reference:24581
Brand Name:Fred EasyPort Defibrillator
Model:
Affected:Multiple Lot numbers
Software version:
Recalling Organisation:InterMed Medical Ltd
71 Apollo Drive
Albany
AUCKLAND 1311
Contact Information:0800 333 444
Manufacturer:Schiller Medical SAS
Issue:It was discovered that electrode material had been caught between a rivet and the ring terminal of the electrode cable leading to no or only intermittent electrical contact between the electrode and the defibrillator.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/5/2019