Detail of Recall

Type of Product:Device
Medsafe Reference:24571
Brand Name:NAV - O-arm Software Upgrade, Version 3.1
Model:Product Numbers: BI-700-00027-xxx and BI-700-00028-xxx, (where XXX denotes system voltage)
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Breakaway Imaging LLC
Issue:The O-arm™ 1000 Imaging System uses energy from the batteries to generate X-rays and move the system. If a charger board is not functioning properly, the batteries will not receive a full recharge of the potential capacity. This can result in the inability of the system to take X-rays, open the gantry, and move the system from one place to another, which may affect the ability to continue use within the operating room until the charger board and/or the batteries are replaced.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/5/2019