Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 24571 |
| Brand Name: | NAV - O-arm Software Upgrade, Version 3.1 |
| Model: | Product Numbers: BI-700-00027-xxx and BI-700-00028-xxx, (where XXX denotes system voltage) |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Breakaway Imaging LLC |
| Issue: | The O-arm™ 1000 Imaging System uses energy from the batteries to generate X-rays and move the system. If a charger board is not functioning properly, the batteries will not receive a full recharge of the potential capacity. This can result in the inability of the system to take X-rays, open the gantry, and move the system from one place to another, which may affect the ability to continue use within the operating room until the charger board and/or the batteries are replaced. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 10/5/2019 |