Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 24550 |
| Brand Name: | ARCHITECT c Systems Mixer |
| Model: | |
| Affected: | List numbers - 09D59-01, 09D59-02, 09D59-03 |
| Software version: | |
| Recalling Organisation: | Abbott Diagnostics Division Ground Floor, Building D 4 Pacific Rise Mt Wellington AUCKLAND 1640 |
| Contact Information: | 0800 656 233 |
| Manufacturer: | Abbott Diagnostics Division |
| Issue: | There is the potential for the mixer blade, attached to the mixer with a screw and nut, to separate from the mixer. The missing mixer blade will result in inadequate mixing of the reaction mixtures leading to potential incorrect results (failed assay calibration, Quality Control (QC) out of range, and/or incorrect patient results). |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 7/5/2019 |