Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:24536
Brand Name:DiaMed GmbH Immunohematology control IH-QC1
Model:
Affected:Catalog number - 009321 - All lot numbers until further notice
Software version:
Recalling Organisation:Bio-Rad New Zealand
Level 30
48 Shortland Street
AUCKLAND 1010
Contact Information:0508 805 500
Manufacturer:Diamed NZ div of DiaMed Australia Pty Ltd
Issue:The manufacturer has been made aware that the anti-D (RH1) 0.05 IU/ml contained in the IH-QC1 has given weaker reactions and sometimes negative results against D positive cells in 2 stage papain technique (set of cells ID-DiaCell IP-IIP-IIIP, ID-DiaPanel P and ID-DiaScreen VP-VIP).
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/5/2019