Type of Product:Device
Medsafe Reference:24531
Brand Name:Therapeutic plasma exchange system
Model:
Affected:Prismaflex devices with software version 8.10
Software version:
Recalling Organisation:Baxter Healthcare Ltd
33 Vestey Drive
Mt Wellington
AUCKLAND
Contact Information:0800 229 837
Manufacturer:Gambro Lundia AB
Issue:The manufacturer has received reports of Prismaflex 8.10 devices with inactive syringe pump during Continuous Renal Replacement Therapy (CRRT) treatment while using Regional Citrate Anticoagulation (RCA). It was determined that there is a potential for the calcium syringe pump to be inactive without the device alarming after a completed change syringe procedure.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/5/2019