Detail of Recall

Type of Product:Device
Medsafe Reference:24524
Brand Name:MITG - Catheter Introducing Sheath, Covidien LLC
Model:MAHURKAR acute catheters
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Covidien LLC
Issue:It has been established that the priming volume values within the instructions for use (IFU) and printed on the MAHURKAR acute catheters are higher than the volumes required to fill each lumen. In some instances, priming volumes for the MAHURKAR acute catheters were established using the maximum calculated volume plus an additional volume to ensure the catheter was fully locked. This approach was chosen to ensure that the lumen is always fully primed, minimizing risk of catheter thrombosis.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/5/2019