Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:24518
Brand Name:Access Family of Immunoassay Systems
Model:
Affected:Access 2, UniCel DxI 600, UniCel DxI 800, UniCel DxC 600i, UniCel,DxC 660i, UniCel DxC 680i, UniCel DxC 860i, and UniCel DxC 880i systems.
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Issue:The manufacturer has identified a potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays following a field action from 2017 (FA-30796), which tested biotin levels up to 100 ng/mL.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/4/2019