| Type of Product: | Device |
| Medsafe Reference: | 24508 |
| Brand Name: | Lights - Ardon |
| Model: | All Getinge VOLISTA light head of model STANDOP and TRIOP manufactured until July 2017 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Getinge Australia Pty Ltd (NZ) Unit D, Ground Floor, Building A, Millenium Centre 600 Great South Rd Ellerslie Auckland 1051 |
| Contact Information: | 09 272 9039 |
| Manufacturer: | Maquet SA |
| Issue: | The manufacturer has identified an issue that might prevent the device from performing as intended. The manufacturer, received complaints regarding light head fork (yokes). It was established that the coating of those forks might show cracks. Therefore, there is a risk that particles may fall and generate infection |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 26/4/2019 |