Detail of Recall

Type of Product:Device
Medsafe Reference:24468
Brand Name:MX40 Telemetry Device
Model:All MX40 Patient Worn Monitors with revision B
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:The manufacturer has identified that some MX40 Patient Worn Monitors may have excessive battery power consumption. For those MX40's with revision B software and fully charged batteries may last approximately 25% shorter than is specified in the MX40 Instructions for Use. This issue affects units operating on both rechargeable and disposable batteries.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/4/2019