Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:24440
Brand Name:Alaris Plus Volumetric Pumps_BD Swiss
Model:All BD Alaris™ VP Pump
Affected:
Software version:
Recalling Organisation:Becton Dickinson Ltd
Unit 4 / 110 Mays Road
Onehunga
Auckland 1061
Contact Information:0800 572 468
Manufacturer:BD Switzerland Sarl
Issue:The manufacturer has identified two issues with the VP LVP pump that results in situations where:
- The upstream occlusion does not generate within the prescribed time period.
- The user faces upstream occlusion when there is no blockage in the infusion path creating nuisance alarms.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/4/2019