Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:24434
Brand Name:(Berchtold Holding GMBH) - I - 12282 - Light, surgical
Model:F 528/F 628 Light Head Covers
Affected:
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:Berchtold GmbH & Co KG
Issue:The manufacturer has become aware that there is a remote likelihood that the Berchtold Chromophare F 528 and F 628 Light Head Cover (Manufactured from September 21st, 2017 to November 13th, 2018) may have an insufficiently assembled component on the back cover. If this component on the Chromophare Light Head Cover disengages, it could potentially fall off the light and cause temporary or reversible injury to HealthCare Providers and/or patients
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/4/2019