Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 24434 |
| Brand Name: | (Berchtold Holding GMBH) - I - 12282 - Light, surgical |
| Model: | F 528/F 628 Light Head Covers |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Stryker New Zealand Ltd 515 Mount Wellington Highway Mount Wellington AUCKLAND 1060 |
| Contact Information: | 0800 787 9287 |
| Manufacturer: | Berchtold GmbH & Co KG |
| Issue: | The manufacturer has become aware that there is a remote likelihood that the Berchtold Chromophare F 528 and F 628 Light Head Cover (Manufactured from September 21st, 2017 to November 13th, 2018) may have an insufficiently assembled component on the back cover. If this component on the Chromophare Light Head Cover disengages, it could potentially fall off the light and cause temporary or reversible injury to HealthCare Providers and/or patients |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 11/4/2019 |