Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 24430 |
| Brand Name: | Mobile Digital X Ray System |
| Model: | MobileDiagnost wDR |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems DMC GmbH |
| Issue: | Unexpected movement may occur while driving the system. The system may exhibit unintended movement which can only be stopped by engaging the emergency stop. In certain circumstances the system may exhibit erratic movement which is stopped by releasing the deadman switch. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 11/4/2019 |