Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:24415
Brand Name:Mobile OR-Table MEERA
Model:
Affected:Material Numbers: 720001B0, 720001F2.
Software version:
Recalling Organisation:Getinge Australia Pty Ltd (NZ)
Unit D, Ground Floor, Building A, Millenium Centre
600 Great South Rd
Ellerslie
Auckland 1051
Contact Information:09 272 9039
Manufacturer:Maquet GmbH
Issue:The manufacturer has received complaints involving a mobile OR-Table MEERA, in which the plastic socket of the IEC 60320-1 socket was pulled out of the metal cage and electronic parts were exposed. There is a risk that a user or third person may suffer an electric shock when coming in contact with the exposed electronic components.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/4/2019
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /