Detail of Recall

Type of Product:Device
Medsafe Reference:24391
Brand Name:Infinity Acute Care System Monitoring - M540
Model:(IACS); sw VG7.0.1; & VG7.0.2 with generation 3 Cockpits
Affected:
Software version:
Recalling Organisation:Draeger New Zealand Ltd
Unit 4
24 Bishop Dunn Place
East Tamaki
Auckland 2013
Contact Information:0800 559 186
Manufacturer:Draegerwerk AG & Co KGaA
Brand Name:INFINITY Patient Monitors & Accessories
Model:
Affected:with sw VG7.0.1; & VG7.0.2.
Software version:
Recalling Organisation:Draeger New Zealand Ltd
Unit 4
24 Bishop Dunn Place
East Tamaki
Auckland 2013
Contact Information:0800 559 186
Manufacturer:Draeger Medical Systems Inc
Issue:The manufacturer has identified that immediately after the monitor has been docked to the Infinity M500 Docking Station and a user discharges a patient from the Infinity M540 Patient Monitor, there is a possibility that the communication between the lACS Cockpit and the Infinity M540 Patient Monitor is not established.
If the communication is not established, the lACS Cockpit will not be able to send any commands to the Infinity M540 Patient Monitor, with the exception of the "audio pause".
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/4/2019