Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:24383
Brand Name:Anaesthesia Work Stations
Model:
Affected:All upgraded Aisys anesthesia devices (i.e. Aisys devices upgraded to software version 11SP01 or 11SP02) and Aisys CS2 anesthesia devices with software version 11SP01 or 11SP02 (GTIN: 840682102292) and Et Control option
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:Datex-Ohmeda Inc
Issue:The manufacturer has become aware that : Aisys CS2 anesthesia devices with software version 11 and Aisys anesthesia devices upgraded to software version 11 with End Tidal Control option (when Et Control auto-exits during Pause Gas procedure agent delivery is off and fresh gas total flow is 10 L/min, resulting in the agent concentration quickly decreasing).
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/4/2019
Hide menus