Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:24377
Brand Name:FilmArray - Blood Culture Identification (BCID) Panel
Model:
Affected:Part No.: RFIT-ASY-0126 and RFIT-ASY-0127
Software version:
Recalling Organisation:Medi'Ray New Zealand Ltd
53-55 Paul Matthews Road
Albany
Auckland 0632
Contact Information:0800 633 472
Manufacturer:Biomerieux SA
Issue:The manufacturer has identified an increased risk of false positive Enterobacteriaceae and Escherichia coli results when the BioFire BCID Panel is used with bioMerieux BACT/ALERT® blood culture bottles. The false positive results are primarily Enterobacteriaceae results with a lower rate of concurrent E. coli results
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/3/2019