Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 24346 |
| Brand Name: | Cios Alpha / Fusion |
| Model: | Cios Alpha |
| Affected: | l |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare GmbH |
| Issue: | The manufacturer has identified that the permanent operation of the Cios Alpha at very high tube outputs results in increased wear of the radiation generation components. When tube voltages of 125 kV are used over a longer period of time, e.g. more than 30 minutes. There is a risk of premature failure of the monoblock and thus a loss of the imaging X-rays. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 22/3/2019 |