Detail of Recall

Type of Product:Device
Medsafe Reference:24340
Brand Name:CBS HS Cryopreservation Straws
Model:
Affected:Leica CM1100; Leica CM1510 S; Leica CM1520; Leica CM1850; Leica CM1850UV; Leica CM1860/UV; Leica CM1950; Leica CM3050 S.
Software version:
Recalling Organisation:Bio-Strategy Ltd
22a William Pickering Drive
North Harbour
Albany
Auckland 0751
Contact Information:09 969 9150
Manufacturer:Cryo Bio Systems - IMV Division
Issue:The manufacturer recognized that some customers are using flammable freezing sprays within the cryostat chamber. Such flammable freezing sprays can potentially ignite and can cause serious injuries. The current labelling does not include any specific warning about usage of flammable freezing sprays.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:15/4/2019