Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:24339
Brand Name:cobas p501/p701
Model:
Affected:Loose screws on the Tray Input Flap
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:Roche Diagnostics GmbH
Issue:Visual inspections of the Tray Input Flap, plate and screws were performed and in some cases found that the flap was no longer attached to the instrument and the 4 M3 threads on the Tray Input Flap were worn out resulting in the screws not working as intended.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/3/2019
Hide menus
Show menus