Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 24339 |
| Brand Name: | cobas p501/p701 |
| Model: | |
| Affected: | Loose screws on the Tray Input Flap |
| Software version: | |
| Recalling Organisation: | Roche Diagnostics (NZ) Ltd 15 Rakino Way Mt Wellington AUCKLAND 1644 |
| Contact Information: | 0800 363 622 |
| Manufacturer: | Roche Diagnostics GmbH |
| Issue: | Visual inspections of the Tray Input Flap, plate and screws were performed and in some cases found that the flap was no longer attached to the instrument and the 4 M3 threads on the Tray Input Flap were worn out resulting in the screws not working as intended. |
| Recall action type: | Correction |
| Recall action: | Product to be exchanged |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 21/3/2019 |