Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:24326
Brand Name:CN - Activa Application Software A610
Model:A610 Deep Brain Stimulation (DBS) Clinician Programmer Application
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Issue:The A610 software is being updated to Version 1.1.3484 which will includes a Charge Density Warning to alert clinicians during programming of settings that exceed 30µC/cm2/phase and provide confirmation to use the selected value.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Healthcare Professional
Date Commenced:20/3/2019