Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 24326 |
| Brand Name: | CN - Activa Application Software A610 |
| Model: | A610 Deep Brain Stimulation (DBS) Clinician Programmer Application |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic Inc |
| Issue: | The A610 software is being updated to Version 1.1.3484 which will includes a Charge Density Warning to alert clinicians during programming of settings that exceed 30µC/cm2/phase and provide confirmation to use the selected value. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Healthcare Professional |
| Date Commenced: | 20/3/2019 |