Detail of Recall

Type of Product:Device
Medsafe Reference:24311
Brand Name:cobas e 801
Model:
Affected:cobas e 801 module, cobas e 801 analytical unit, ProCell II M
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:Roche Diagnostics GmbH
Issue:On affected cobas e 801 modules/analytical units, the ProCell II M flowpath is covered by a biofilm with ProCell II M resistant bacteria. If ProCell II M inside the flowpath is not exchanged regularly, bacteria can degrade the components of ProCell II M resulting in a reduced ECL signal generating capacity.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/3/2019