Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 24308 |
| Brand Name: | Infinity Acute Care System Monitoring - M540 |
| Model: | |
| Affected: | with sw VG7.0.1 |
| Software version: | |
| Recalling Organisation: | Draeger New Zealand Ltd Unit 4 24 Bishop Dunn Place East Tamaki Auckland 2013 |
| Contact Information: | 0800 559 186 |
| Manufacturer: | Draegerwerk AG & Co KGaA |
| Brand Name: | INFINITY Patient Monitors & Accessories |
| Model: | |
| Affected: | with sw VG7.0.1 |
| Software version: | |
| Recalling Organisation: | Draeger New Zealand Ltd Unit 4 24 Bishop Dunn Place East Tamaki Auckland 2013 |
| Contact Information: | 0800 559 186 |
| Manufacturer: | Draeger Medical Systems Inc |
| Issue: | The manufacturer has identified that when the ECG cable plug is not properly connected, the Infinity M540 Patient Monitor may reboot. If the situation is not corrected and the reboot occurs 3 times in 10 minutes or less, the Infinity M540 Patient Monitor will enter a fail-state, resulting in a resetting to factory defaults, losing patient settings and locally stored patient data. The reboot and the fail state are indicated by visual and audio alarms. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 14/3/2019 |