Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 24289 |
| Brand Name: | CareFusion_Blood Warmer Cartridges |
| Model: | enFlow Disposable Sterile Cartridge with IV Extension Set |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Clinical & Regulatory Services Limited 30 Fulford Road Havelock North Hastings 4130 |
| Contact Information: | (06) 844 4490 |
| Manufacturer: | CareFusion USA |
| Issue: | The manufacturer has identified that uncoated aluminium plates in fluid warming systems can lead to a risk of administering potentially harmful concentrations of aluminium, when using balanced electrolyte solutions. The manufacturer is commencing a supplier-initiated global recall involving the enFlow Disposable Cartridge used with the enFlow® fluid warming system enFlow®. This medical device recall supersedes the Product Safety Alert dated 13 March 2019. |
| Recall action type: | Recall |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 14/3/2019 |