Detail of Recall

Type of Product:Device
Medsafe Reference:24288
Brand Name:Philips MultiDiagnost-Eleva system
Model:MultiDiagnost Eleva, MultiDiagnost Eleva with Flat Detector, OmniDiagnost Eleva, OmniDiagnost Classic
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:During service of two MultiDiagnost Eleva systems, it was noticed that a locking plate was missing. In one of these cases, the shaft was still properly positioned after the system had been in use for 9 years. In the other case, involving a system that had been in use for 5 years, the shaft was no longer properly positioned but was still holding the actuator.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/3/2019