Detail of Recall

Type of Product:Device
Medsafe Reference:24276
Brand Name:FilmArray Gastrointestinal (GI) Panel
Model:
Affected:Part numbers: RFIT-ASY-0104 & RFIT-ASY-0116
Software version:
Recalling Organisation:Medi'Ray New Zealand Ltd
53-55 Paul Matthews Road
Albany
Auckland 0632
Contact Information:0800 633 472
Manufacturer:BioMerieux Inc
Issue:The manufacturer has identified elevated rates of false positive results for Campylobacter and Cryptosporidium while using the BioFire GI Panel. Only the BioFire GI Panel kits with expiration dates ranging from 10/23/2019 - 02/20/2020 are affected.
This is an expansion of previous Urgent Product Correction notice Ref #24276
The manufacturer has identified an increased risk of false positive results for Campylobacter and Cryptosporidium when using the BioFire GI Panel with expiration dates of 23 October 2019 onward. An investigation has determined that the false positive results are due to non-specific PCR amplification in the test, not from contamination with these two organisms.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/3/2019