Detail of Recall

Type of Product:Device
Medsafe Reference:24272
Brand Name:Datascope Balloon pump
Model:CARDIOSAVE Hybrid IABP, CARDIOSAVE Rescue IABP
Affected:
Software version:
Recalling Organisation:Getinge Australia Pty Ltd (NZ)
Unit D, Ground Floor, Building A, Millenium Centre
600 Great South Rd
Ellerslie
Auckland 1051
Contact Information:09 272 9039
Manufacturer:Datascope Corporation
Issue:The manufacturer has received complaints involving the Cardiosave IABPs regarding the use of certain Intra-Aortic Balloons (IABs) at altitudes above 975 Meters. The Cardiosave may not successfully complete the autofill process required to initiate pumping. This failure may result in either interruption of therapy upon the first maintenance autofill or the inability to start therapy.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/3/2019