Type of Product:Device
Medsafe Reference:24252
Brand Name:Angiographic X-Ray
Model:
Affected:Allura Xper FD20 - Bent Pedals of Footswitch
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:The manufacturer has identified that if a Footswitch is frequently used on an anti-fatigue mat, on a non-flat surface or in the pedestal, the footswitch pedals may get bent, which might lead to an intermittent or continuous inability of making live fluoroscopy images or exposures.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:4/3/2019