Detail of Recall

Type of Product:Device
Medsafe Reference:24239
Brand Name:Elecsys® FT4 III assays.
Model:
Affected:Elecsys FT4 III (100 tests) - MODULAR ANALYTICS E170, cobas e 411, cobas e 601, cobas e 602, Elecsys FT4 III E2G (300 tests) - cobas e 801
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:Roche Diagnostics GmbH
Issue:The manufacturer has identified a reduced biotin interference threshold for the recently launched ElecsysR FT4 III assays. The described threshold in the method sheet is 409 nmol/L or 100 ng/mL of biotin in a sample. However, for the following reagent lots the threshold is 81.8 nmol/L or 20 ng/mL. In patient samples with biotin concentrations of 266 nmol/L or 65 ng/mL, FT4 recovery could lead to a difference of >20%.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:27/2/2019