Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 24187 |
| Brand Name: | Sara 3000, Plus, Lite, Flex, Combilizer. Maxi Twin, Compact, Move, 500. Maxi Sky 440, 600, 1000, Maxi Sky 2, Maxi Sky 2 Plus, Maxi Lite, Walker, Minstrel, Tenor; Miranti; Voyager; Bolero |
| Model: | SN P0402544 to SN P0485889 (1 455 units manufactured in total) |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Arjo New Zealand Limited 34 Vestey Drive Mt Wellington AUCKLAND 1060 |
| Contact Information: | 0800 000 151 |
| Manufacturer: | ArjoHuntleigh AB |
| Issue: | It has been identified that the affected lifts operating with software version 0.3, 0.4 or 0.5, without a heatsink component applied to the PCB, can contribute to the premature wear of this part and prevent the user from operating the lift as outlined in the Instructions for Use (IFU), in particular utilization of raising/lowering or legs opening/closing function. The potential for this malfunction is caused by a fault in the PCB software. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 6/7/2020 |