Detail of Recall

Type of Product:Device
Medsafe Reference:24170
Brand Name:Stryker Spine - I - 44759 - Fixation system implantation kit, internal
Model:XIA 3 FRENCH BENDER, XIA 3 DEGENERATIVE INSTRUMENTS, XIA 3 COMPLEX ROD & MAC INST
Affected:
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:Stryker Spine
Issue:The Xia 3 Degen Tray and the French Bender listed above did not successfully pass validation for a steam sterilisation cycle of 134°C for 3 minutes as the French Bender instrument contained within the tray did not meet the required sterility assurance level (SAL) of 10-6. Furthermore, since the Xia 3 Rod and Cross connector Tray has the same device design and configuration with respect to the French Bender instrument, it has been determined that it would also not successfully pass validation.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:13/2/2019