Detail of Recall

Type of Product:Device
Medsafe Reference:24153
Brand Name:ARCHITECT Estradiol Reagent Kit
Model:
Affected:ARCHITECT Estradiol Reagent Kit - 7K72-20 All Lot numbers, Alinity i Estradiol Reagent Kit - 7P5020 All Lot numbers
Software version:
Recalling Organisation:Abbott Australasia Pty Ltd (NZ)
Abbott Laboratories NZ
4 Pacific Rise, Mt Wellington
Auckland 1060
Contact Information:61 2 93849717
Manufacturer:Abbott Diagnostics
Issue:The manufacturer has confirmed that the drug Mifepristone may interfere/crossreact with the ARCHITECT Estradiol assay (LN 7K72) and Alinity i Estradiol assay (LN 07P50) leading to falsely elevated Estradiol results.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/2/2019