Detail of Recall

Type of Product:Device
Medsafe Reference:24152
Brand Name:Volcano - Imaging Systems
Model:Volcano s5i, CORE and CORE MOBILE systems
Affected:
Software version:
Recalling Organisation:Device Technologies New Zealand
47 Arrenway Drive
Albany
AUCKLAND 0632
Contact Information:09 913 2000
Manufacturer:Volcano Corporation
Issue:The version 2.5 FFR/iFR software is designed to calibrate the aortic output of the Hemodynamic Monitoring system (HMS). If the HMS adheres to commonly used pressure calibration values, then the interoperability issue will not have any impact. However, if the system does not adhere to commonly used values , the Volcano Pa(Ao) and the HMS AO may be misaligned by up to +/- 5 mm Hg. This discrepancy may lead to confusion
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/2/2019