Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 24142 |
| Brand Name: | Physio-Control Lifepak 15 |
| Model: | LIFEPAK 15 Monitor/Defibrillator |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Stryker New Zealand Ltd 515 Mount Wellington Highway Mount Wellington AUCKLAND 1060 |
| Contact Information: | 0800 787 9287 |
| Manufacturer: | Physio-Control Corporation |
| Issue: | The manufacturer has become aware that certain LIFEPAK® 15 Monitor/Defibrillators were reported to experience a lock-up condition after a defibrillation shock was delivered. This condition is defined as a blank monitor display with LED lights on, indicating power to the device, but no response in the keypad and device functions. A device in this condition has the potential to delay delivery of therapy, and this delay in therapy has the potential to result in serious injury or death. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 4/2/2019 |