Detail of Recall

Type of Product:Device
Medsafe Reference:24123
Brand Name:BioPlex 2200 ANA Screen Control Set
Model:
Affected:Lot No. 50475
Software version:
Recalling Organisation:Bio-Rad New Zealand
Level 30
48 Shortland Street
AUCKLAND 1010
Contact Information:0508 805 500
Manufacturer:Bio-Rad (France)
Issue:The manufacturer has identified that there is an error on the printed value assignment insert for the BioPlex 2200 ANA Screen Control (Catalogue No. 663-1131, UDI 00847865000055, Lot No. 50475). Product Support has determined that the mean values of the negative control for SS-A 60 and Centromere B were incorrectly listed as 0.2 Al on the printed value assignment insert for the BioPlex 2200 ANA Screen Control Set. The correct value for the negative control of SS-A 60 is 0.4 Al and the correct value of Centromere B is 0.3 Al. The mean values that were entered into your BioPlex 2200 System via the control assignment CD were correct and no further action is needed.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/1/2019