Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:24096
Brand Name:Catheter, cardiac, balloon, thermal dilution
Model:
Affected:Lot number: 61321241
Software version:
Recalling Organisation:Edwards Lifesciences (New Zealand) Limited
171 Morrin Road
St John
Auckland 1072
Contact Information:021 313 234
Manufacturer:Edwards Lifesciences LLC
Issue:The manufacturer has become aware of an issue that involves Swan Ganz catheters with incorrect lumen assembly, which causes reversal of the lumens. If the lumens are reversed the clinician may note reverse PA and CVP pressure values and waveforms. It is possible for unintended treatment due to inaccurate values if the reversal in waveform is not noticed.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/1/2019