Detail of Recall

Type of Product:Device
Medsafe Reference:24076
Brand Name:Efficia CM Series
Model:Battery Management Software Release
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:The manufacturer has received a report in which a Philips Efficia Monitor (CM100) with lithium ion batteries that had exceeded its battery life expectancy and overheated and ignited. These batteries should be replaced when the number of charge-discharge cycles exceed 300 cycles or when the remaining battery capacity is below 80% of that of a new battery, whichever comes first.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:15/1/2019