Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 24076 |
| Brand Name: | Efficia CM Series |
| Model: | Battery Management Software Release |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems |
| Issue: | The manufacturer has received a report in which a Philips Efficia Monitor (CM100) with lithium ion batteries that had exceeded its battery life expectancy and overheated and ignited. These batteries should be replaced when the number of charge-discharge cycles exceed 300 cycles or when the remaining battery capacity is below 80% of that of a new battery, whichever comes first. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 15/1/2019 |