Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 24073 |
| Brand Name: | CRDM - Carelink Encore Programmer Model 29901 |
| Model: | 2090 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic Inc |
| Brand Name: | Medtronic model 2090 programmer |
| Model: | 29901 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic Inc |
| Issue: | Software update Model SW022 version 8.1 is required to ensure that the correct remaining longevity estimate is displayed at all times during the life of the device. Without the software update, the displayed value during this period may remain static and reflect an over-estimation of the remaining longevity. Once the RRT voltage threshold is reached the correct remaining longevity estimate is once again displayed. Per labelling, the time from RRT trigger to End of Service (EOS) is 180 days |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 14/1/2019 |