Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 24072 |
| Brand Name: | Stryker Spine Inc. - IIb - 38161 - Spinal, cage |
| Model: | Tritanium PL Cage Surgical Techniques |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Stryker New Zealand Ltd 515 Mount Wellington Highway Mount Wellington AUCKLAND 1060 |
| Contact Information: | 0800 787 9287 |
| Manufacturer: | Stryker Spine |
| Issue: | The manufacturer has received product inquiries regarding Tritanium PL cage fractures occurring intraoperatively and post-operatively. Therefore, the manufacturer has made a decision to update the Surgical Technique to caution against misuse, which includes the use of surgical techniques that fall outside the approved instructions for use. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 14/1/2019 |