Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 24068 |
| Brand Name: | Xper Flex Cardio Physiomonitoring System |
| Model: | Service Numbers: 453564241901, 453564241911, 453564483321, 453564483331, 453564621791, 453564621801, 453564669081 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Invivo, a division of Philips Medical Systems |
| Issue: | Under certain conditions, real time waveforms may not be accurately plotted or displayed. This problem is related to the display of waveforms only, and all numeric data including invasive blood pressure (IBP), NIBP, respiratory rate, and end tidal CO2 are correct. Should this occur during the plotting of an ECG QRS complex on the ECG channel, the R wave may be shortened or missing or the QRS complex may appear slightly broader. A pressure or other phasic waveform may appear to have a "notch" or a "flattened" appearance. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 11/1/2019 |