Type of Product:Device
Medsafe Reference:24024
Brand Name:DEAL COPY - Olympus - Optical forceps biopsy graspers, endoscopic forceps (Germany)
Model:
Affected:Optical Forceps, Instructions for Use
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Olympus Winter & Ibe GmbH
Brand Name:DEAL COPY - Olympus - Bladder stone punches (Germany)
Model:
Affected:Optical Forceps, Instructions for Use
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Olympus Winter & Ibe GmbH
Issue:The manufacturer has been made aware that application of an excessive force to the handle can damage the taut wire inside the instrument making the jaws inoperable.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/12/2018