Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 24010 |
| Brand Name: | Instrument: XN-10 / 20 |
| Model: | XN-10 Serial numbers between 35612 ¡V 36264 and XN-20 Serial numbers: 14090 - 14151 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Cenexi SAS 52 rue Marcel et Jacques Gaucher Fontenay-sous-Bois 94120 France |
| Contact Information: | |
| Manufacturer: | Sysmex Corporation |
| Issue: | The manufacturer has identified that some XN instruments have incomplete manufacturing of the wiring cord for the FCM detector block. If the cord is moved or touched it may result in a shift to the right of the WDF scattergram which will impact on flagging of the WBC differential parameters. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 18/12/2018 |