Detail of Recall

Type of Product:Device
Medsafe Reference:24006
Brand Name:Dimension Vista® 500 Intelligent Lab System
Model:
Affected:Millipore Progard Pretreatment Pack - Lot numbers - F7MA56608, F7MA50339, F7NA62488, F8AA93567, F8HA64477, F8AA13966, F8HA79819, F8BA20457, F8BA35774, F8DA59251, F8HA91305, F8JA97876, F8MA06374
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:The manufacturer has identified an issue with Millipore Water Purification Module (WPM) Progard Pretreatment Packs when used with the Dimension Vista® Triglyceride assay. An investigation identified the potential for patient and quality control (QC) triglyceride results to be erroneously elevated as a result of contamination introduced by the Millipore Progard Pretreatment Pack's filter fibers. A positive bias of up to 1.38 mmol/L has been observed. The potential for falsely elevated results could occur when processing the triglyceride assay immediately following the change of the Progard Pretreatment Pack in the Millipore WPM.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/12/2018