| Type of Product: | Device |
| Medsafe Reference: | 23978 |
| Brand Name: | MOSAIQ |
| Model: | Using release version 1.0 and higher and Multi-ACCESS using release version 6.1 and higher. |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Elekta Pty Limited - New Zealand Branch Suite 302, Geyser Building, 100 Parnell Road Parnell AUCKLAND 1052 |
| Contact Information: | 0273-544-793 |
| Manufacturer: | Elekta Inc |
| Issue: | If there are no Wedge IDs included in the DICOM RT PLAN sent from the TPS, the field in MOSAIQ will be created with no wedge. If not discovered, treatment of a field without the planned wedge will receive more dose than the treatment plan indicates. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 13/12/2018 |