Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 23968 |
| Brand Name: | Stryker Communications - I - 32245 - Medical monitor/light mount |
| Model: | |
| Affected: | Item numbers - 0682001567I, 0682001269I, 0682001267, 0682001268, 0682001566, 0682001570I, 0682400142 |
| Software version: | |
| Recalling Organisation: | Stryker New Zealand Ltd 515 Mount Wellington Highway Mount Wellington AUCKLAND 1060 |
| Contact Information: | 0800 787 9287 |
| Manufacturer: | Stryker Communications Inc |
| Issue: | The manufacturer has become aware that there is a remote likelihood that the EDS Light Suspension, Central Axis (Manufactured from March 1, 2014 to October 31, 2015) may have an insufficiently assembled joint in the suspension that is not visible to the user. The EDS systems have redundant connections throughout the suspensions. If the EDS Light Suspension, Central Axis falls, it could potentially cause serious injuries. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 11/12/2018 |