Detail of Recall

Type of Product:Device
Medsafe Reference:23946
Brand Name:Biograph mCT Hoffman Estates
Model:
Affected:Biograph Horizon - Oversized CT Circuit Breaker
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Medical Solutions USA Inc
Issue:The manufacturer has become aware of an oversized (70amp) breaker in the PDCC (power distribution computer cabinet). The 70amp oversized breaker could, in an unforeseen situation where the current draw exceeds 60amps, but is not enough to trip the 70amp breaker, cause a possible distinct burning smell and the system may cease all functionality.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/12/2018