Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 23937 |
| Brand Name: | Nuclear Medicine System, Gamma Camera, Stationary |
| Model: | Discovery NM 530c and Ventri Nuclear Medicine system |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Medical Systems Functional Imaging Haifa |
| Issue: | The manufacturer has identified that a small number of Nuclear Medicine systems are operating without currently updated software. In a recent event, a patient's finger was injured during the Unload process while the automatic Unload feature was being used; the software on the system was not current |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 10/12/2018 |