Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 23922 |
| Brand Name: | MOSAIQ |
| Model: | MOSAIQ release version 2.64 SP9 and higher, and Varian TrueBeam or Varian 4D or Siemens RTT |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Elekta Pty Limited - New Zealand Branch Suite 302, Geyser Building, 100 Parnell Road Parnell AUCKLAND 1052 |
| Contact Information: | 0273-544-793 |
| Manufacturer: | Elekta Inc |
| Issue: | The manufacturer has identified that if a Setup field (CT, kV, MV or MVCT) is created by COPYING a treatment field or another setup field, then the copied field will have the same isocenter as the parent and that isocenter may not be the correct isocenter for imaging when it is sent to the machine. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 3/12/2018 |