Detail of Recall

Type of Product:Device
Medsafe Reference:23922
Brand Name:MOSAIQ
Model:MOSAIQ release version 2.64 SP9 and higher, and Varian TrueBeam or Varian 4D or Siemens RTT
Affected:
Software version:
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Elekta Inc
Issue:The manufacturer has identified that if a Setup field (CT, kV, MV or MVCT) is created by COPYING a treatment field or another setup field, then the copied field will have the same isocenter as the parent and that isocenter may not be the correct isocenter for imaging when it is sent to the machine.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/12/2018