Detail of Recall

Type of Product:Device
Medsafe Reference:23920
Brand Name:cobas c111 Chemistry Analysers
Model:All cobas c111 Chemistry Analysers (with and without ISE)
Affected:
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:Roche Diagnostics GmbH
Issue:The results of an investigation show that a reagent carryover to CREP2 or to the Triglycerides test (TRIGL) might occur if a preceding test containing dextran sulfate (DS) (e.g. HDL-Cholesterol Gen.4 [HDLC4]) is pipetted first and is then followed by Cholesterol Gen.2 (CHOL2). The manufacturer requires the installation of extra wash cycles (EWCs) after CHOL2 as this prevents any observed high flyer results. Secondary carry-over to TRIGL or CRE2 occurs only if a dextran sulfate containing assay is performed before CHOL2. Hence these EWCs are only mandatory in case one of the following tests is also installed on the analyzer: HDLC4, HDLC3 or LDLC3.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/12/2018