| Type of Product: | Device |
| Medsafe Reference: | 23896 |
| Brand Name: | syngo MR |
| Model: | Syngo.plaza version VB30A_HF01 and previous software versions |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare GmbH |
| Issue: | The manufacturer has identified that the display orientation labels may not be displayed properly, this could occur when images have been loaded and the user performs Rotate and Flip operations together and saves changes. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 27/11/2018 |